ConnectorTest International 4097A, 4087B, 4077A
ADVANCED TEST SYSTEMS








The 4077A, 4087B and 4097A systems STAND ALONE in the commercial ATE marketplace by OUTPERFORMING ALL OTHER COMMERCIALLY ATE systems with their unique combination of superior: 

1)  Speed 

2)  Test Function

3)  Ease-of-Operation 

4)  Portability

5)  Serviceability

6)  Return on Investment (ROI)

At one-third the cost of some other ATE systems, these advanced systems from ConnectorTest can achieve a return on investment (ROI) in as little as four months. 

MODEL 4087B & 4097A PERFORMANCE SPECIFICATIONS 

TEST METHOD:

One pin tested against all other pins tied to the connector's Ground shell.

CAPACITANCE:

1.0 pF - 1.0 uF, 1% typical; 2% max / 1KHz test frequency

 - Test Capacitance at DC Bias Voltages from 0 - 750 VDC

INSULATION RESISTANCE (IR):

    5 - 100,000 Meg Ohms (100 Gig Ohm), 2% (200 Gig Ohm Option)

    0 - 1500 VDC Test Voltage; 0.1%

    0.1 - 650 second Programmable Dwell Time Setting

DIELECTRIC WITHSTANDING VOLTAGE ( DWV / Hi-Pot ):

    DC:

    0 - 1500 VDC Test Voltage; 0.1%

    0 - 2500 VDC Option

    0 - 1000 uA Fail Current; 2%
 

    AC:

    0 - 1800 VAC Test Voltage, 60 Hz; 2%

    0 - 2300 VAC Option

    0 - 2000 uA Programmable Fail Current

    0.1 - 650 Second Programmable Dwell

    AC Ramp Options:  

    a) 500 VAC / Sec. (Space Station Requirement)

    b) 200 VAC / Cycle

 

ZENER DIODE / TVS / MOV DEVICES:

    0 - 750 VDC Test Voltage; Bi-Polar / Uni-Polar       

    1 - 10 mA Test Current

    4.8 - 750 VDC Breakdown Voltage

    Selectable Reverse Leakage Limits:  1nA - 10,000 nA; 1uA - 1000uA

SHORTED PIN TEST:

Verify intentionally shorted pins, either to Ground or to another pin.


GROUND RING TEST:

Verify presence of Ground ring on connector / cable shell.  Can be disconnected from test to identify pins shorted to one another.


CABLE CONTINUITY  /  MODEL 4097A ONLY:

    Measure up to 160 Wires (320 terminations); 

    Resistance Measurements: 0.01 Ohms - 4000 Ohm measurements

    Test Current Selectable: 1, 10, 100 mA


CONTACT RESISTANCE:

    Measure 0.004 Ohms - 4.00 Ohms @ 100 mA

    Resolution = 4 milliohms

    Standard on 4097A; Optional on 4087B


AC POWER REQUIREMENTS:

    105 - 125 VAC; 50 - 60 Hz

    350 mA, AC Input Current

    Europe 230 VAC; 50Hz Option

SIZE:

    4087B & 4097A  =  14" W x 14" H x 19" D

    Weight = 45 lbs

MODEL 4077A PERFORMANCE SPECIFICATIONS

TEST METHOD:  

Test clusters of non-adjacent pins; 8 test points available.

INSULATION RESISTANCE:

    5 - 100,000 Meg Ohm; 2%

    0 - 1500 VDC Test Voltage; 0.1%

    0.1 - 650 Second, Programmable Dwell Time

DIELECTRIC WITHSTANDING VOLTAGE:

DC:

    0 - 1500 VDC Test Voltage; 0.1%

    0 - 2500 VDC Option

    0 - 1000 uA Fail Current; 2%

    0.1 - 650 Second Programmable Dwell

AC:

    0 - 1800 VAC, 60 Hz; 2%

    0 - 2300 VAC Option

    0 - 2000 uA Fail Current; 2%

    0.1 - 650 Second Programmable Dwell

    Ramp Options:  

    a) 500 VAC / Sec. 

    b) 200 VAC / Cycle

SIZE:

    4077A = 12.5" W x 8.5" H x 17.5" D

    Weight = 15 Lbs

SYSTEM OPTIONS:


17" or 19" LCD Touchscreen 

Bar-Code Scanner (Saves time by scanning long part numbers, job / lot numbers, etc. into menu text field.)

Temperature / Humidity / Barometric Pressure Instrumentation; displays measurments on Instrument's display, printed documents and Main menu.

UPS (Uninterruptible Power Supply); 500 VA. True sinewave power supply; (Recommended)

Model 4087B can be configured for Contact Resistance measurements.

200,000 Meg Ohm (200 Gig Ohm) Insulation Resistance measurements.

0 - 2500 VDC and 0 - 2300 VAC Test Voltages


| home | about | specifications | features | screens | contact |

Copyright © 1996-2014, ConnectorTest International (TM)
P.O. Box 661315 Arcadia, CA 91066-1315  USA
Tel. 626-445-6200 